UltraNuclease, also known as non-restrictive endonuclease, broad-spectrum nuclease, is a non-specific endonuclease derived from Serratia Marcescen, which hydrolyzes internal phosphodiester bonds between any nucleotides in nucleic acids to produce 5'-monophosphate oligonucleotides of 2-5 bases in length. It can degrade DNA and RNA in various forms (double-stranded, single-stranded, linear, circular, native or denatured) under a very broad range of conditions (6 M Urea, 0.1 M Guanidine HCl, 0.4% Triton X-100, 0.1% SDS, 1 mM EDTA, 1 mM PMSF) and is widely used to remove nucleic acids from biological products. UltraNuclease can also be removed by corresponding methods subsequently.
This kit uses the principle of double-antibody sandwich enzyme-linked immunosorbent assay (sandwich ELISA) to detect the residues of denatured and non-denatured UltraNuclease. First, coat the microplate with anti-UltraNuclease rabbit polyclonal antibody to form a solid-phase antibody. Second, add UltraNuclease standard and test sample to the solid-phase antibody microplate, then add biotin labeled Anti-UltraNuclease polyclonal antibody, and finally add horseradish peroxidase-labeled streptavidin (SA-HRP) to form an antibody + antigen + antibody-Biotin + SA-HRP complex. Subsequently, add TMB substrate into the complex to observe color reaction after washing the complex. TMB is converted into blue under the catalysis of HRP enzyme and finally converted into yellow in the presence of acid, and the shade of color is positively correlated with the amount of UltraNuclease in the sample.
The detection range of this kit is 0.047-3 ng/mL; the lower detection limit is 23.5 pg/mL.
Sensitivity |
23 pg/mL (range 0.047~3 ng/mL) |
Assay Time |
<4 hours |
Assay Principle |
Two-site immunoenzymetric assay |
Signal Amplification |
Biotin-Streptavidin system |
Detection Wavelength |
450nm |
Components No. |
Name |
36718ES59 |
36718-A |
ELISA Microplates |
1 Plate |
36718-B |
Detection Antibody:Biotin-conjugated Rabbit Anti-UltraNuclease Antibodies |
35 μL (0.2 mg/mL) |
36718-C |
Standard: UltraNuclease |
1 vial (500 ng/mL) |
36718-D |
HRP-conjugated Streptavidin |
10 μL |
36718-E |
Dilution Buffer 1 |
45 mL |
36718-F |
20×Wash Buffer |
50 mL |
36718-G |
Dilution Buffer 2 |
30 mL |
36718-H |
TMB |
15 mL |
36718-I |
Sealing Film |
5 pieces |
* The stop solution is not provided in this kit, it need to be prepared by user. 1M HCl can be used as the stop solution.
This product should be stored at 2°C ~8°C. Unopened product is valid for one year. Once the reagent is opened, it is valid for half a year. Never freeze this product.
*1. The diluted washing buffer and dilution buffer can be stored at 2°C ~8°C for 1 week.
*2. Standard 36701-C is liquid and stored at 2°C ~8°C for 1 year.
*3. Prepared detection antibody and stop buffer can be stored at 2°C ~8°C for 1 month.
Figures 1. The specificity of UltraNuclease ELISA kit was shown in the figure above
8 mimic samples were estimated for the non-specific reaction with the antibodies in the kit. There was no significant difference between the 8 samples and the negative sample. (*: Significant difference from Negative sample)
A)Repeatability
Concentration (ng/mL) |
6 duplicated data in the same plate |
CV |
|||||
3 |
2.412 |
2.390 |
2.323 |
2.367 |
2.476 |
2.291 |
2.53% |
1.5 |
1.780 |
1.883 |
1.884 |
1.908 |
1.876 |
1.883 |
2.19% |
0.75 |
1.123 |
1.215 |
1.166 |
1.18 |
1.140 |
1.109 |
3.12% |
0.375 |
0.691 |
0.706 |
0.708 |
0.722 |
0.722 |
0.679 |
2.21% |
0.1875 |
0.449 |
0.462 |
0.453 |
0.460 |
0.447 |
0.433 |
2.13% |
0.094 |
0.306 |
0.308 |
0.304 |
0.311 |
0.295 |
0.302 |
1.65% |
0.047 |
0.242 |
0.241 |
0.240 |
0.241 |
0.256 |
0.240 |
2.37% |
0 |
0.183 |
0.192 |
0.189 |
0.187 |
0.189 |
0.189 |
1.43% |
B)Intermediate precision
Intermediate Precision |
|||
Conc. |
1.5 ng/mL |
0.375 ng/mL |
0.094 ng/mL |
Analyst 1 |
1.388 |
0.382 |
0.079 |
Analyst 2 |
1.458 |
0.363 |
0.068 |
Analyst 3 |
1.318 |
0.370 |
0.063 |
Average |
1.388 |
0.372 |
0.070 |
SD |
0.070 |
0.010 |
0.008 |
CV |
5.04% |
2.59% |
11.69% |
Figures 2. The precision of UCF.ME™ UltraNuclease ELISA was assayed as the above scheme
6 duplicated were tested at each concentration on the standard curve, and all the CV% were lower than 5% (Figure 2A). 3 analysts conduct the UCF.ME™ UltraNuclease ELISA experiment and the CV% were lower than 15% (Figure 2B).
A)
B)
Concentration |
OD450 |
Result |
||||
ng/mL |
Repeat 1 |
Repeat 2 |
Repeat 3 |
Average |
Conc. |
Recovery% |
3 |
2.271 |
2.222 |
2.287 |
2.260 |
2.950 |
98% |
1.5 |
1.495 |
1.488 |
1.514 |
1.499 |
1.540 |
103% |
0.750 |
0.868 |
0.856 |
0.890 |
0.871 |
0.735 |
98% |
0.375 |
0.537 |
0.525 |
0.522 |
0.528 |
0.358 |
95% |
0.188 |
0.373 |
0.369 |
0.389 |
0.377 |
0.201 |
107% |
0.093 |
0.261 |
0.255 |
0.258 |
0.258 |
0.082 |
88% |
0 |
0.184 |
0.195 |
0.187 |
0.189 |
0 |
0 |
Figures 3. The accuracy of UCF.ME™ UltraNuclease ELISA was assayed as the following scheme
UltraNuclease was spiked in 5 μg/mL cell lysis solutions (HEK293, E.coli, CHO, and Vero) independently with a final spiked conc. of 1.5 ng/mL. The accuracy in different cell samples showed excellent than other vendor(Figure 3A). UltraNuclease was spiked in an actual sample (provided by the client) and is serial 2 folds diluted from 3ng/mL to 0.093ng/mL. The spike recovery was all between 80%~110% (Figure 3B).